6-5
BP1 Setup Factory Defaults
BP1 SETUP ADULT PED NEO DESCRIPTION
ALARMS OFF OFF OFF Enables BP1 alarm
ALARM LEVEL 2 2 2 Sets alarm priority level (1, 2, or 3)
LATCH ALARMS OFF OFF OFF Latches the BP1 alarm until reset
SYS HI LIMIT 170 130 100 Systolic high alarm limit setting
SYS LO LIMIT 90 65 50 Systolic low alarm limit setting
DIA HI LIMIT 100 90 70 Diastolic high alarm limit setting
DIA LO LIMIT 50 40 30 Diastolic low alarm limit setting
MN HI LIMIT 120 100 80 Mean high alarm limit setting
MN LO LIMIT 60 50 35 Mean low alarm limit setting
PRAM AVAIL ON ON ON BP1 enable or disable
PRAM ON / OFF OFF OFF OFF BP1 standby
RCD ON ALARM WF WF WF Record when BP1 alarm enabled
SIZE 120 120 120 BP1 waveform size
SCALE OFF OFF OFF BP1 scale enable
DISPD VALUES S/D S/D S/D BP1 numeric display format
SITE OFF OFF OFF BP1 catheter site
PRAM COLOR RED RED RED Sets on-screen color of parameter