12-6
Operating Pressure: 430 to 795 mmHg
Maximum Change Rate in Pressure: 12.4 mmHg/min
Operating Altitude: -1,250 to 15,000 ft.
Maximum Change Rate in Altitude: 500 ft./min
Relative Humidity: 10 to 95%
Ambient CO2: 0 to 700 ppm
Flow Rate: 50 ml.min ±7.5 ml/min
Resp Range: 0 to 150 bpm
BTPS: 37° C, 47 mmHg
System Response Time: 2.9 seconds
Rise Time Adult and Neonatal: 190 msec max
Delay Time: 2.7 seconds typical
CO
2
Report Resolution: 1 mmHg
Calibration Interval: 4000 operating hours or I year
Power:
1.5 Watts Normal Operation
2.5 Watts In AutoZero Mode
0.9 Watts MediCO2 Standby Mode
4.7 Watts Initialization Mode
Accuracy:
TIME CO2 CONCENTRATION* ACCURACY**
0-20 minutes 0-38 mmHg
39-99 mmHg
±4 mmHg
±12% of reading
20 minutes and up 0-38 mmHg
39-99 mmHg
±2 mmHg
±5% of reading + 0.08%
of every 1 mmHg above
40 mmHg
* At sea level.
** Accuracy applies for breath rates of up to 80 bpm. For breath rates above 80 bpm,
accuracy complies with EN 864/ISO 9918 (4 mmHg or 12% of reading whichever is
greater).
The accuracy is maintained to within 4% for the following Vol. % gas mixtures:
CO
2
N
2
O
2
N
2
O H
2
O Anesthetic Agents
0-13 0-97.5 0-100 0-80 Dry-Saturated According to EN864
Leak Tightness: The leak rate of the module flow system is less than 40 mBar per
minute when a 30% vacuum is invoked on the flow system.
THERMAL RECORDER
Channels: Any 1 or 2 waveforms, user selectable
Resolution: 8 dots/mm vertical, 32 dots/mm horizontal
Paper: 40 mm grid; 50 mm width
Sweep Speed: 6.25, 12.5, 25 mm/sec; 1 mm/sec Trend
Manual Record Duration: 16 seconds, 12 seconds memory
Annotation: Time, date, parameter values, scales, source, speed
Weight: 1.15 lb (0.52 kg) with full roll of paper (Prism Only)