Vista de página 68
3-28
5. To remove temporary display, press the PAGE HOME key.
To display OxyCRG in Trace 2:
1. Monitor ECG, Respiration, and SpO2.
2. Press the SYSTEM SETUP key.
CLEAR
PT
DATA
STORE
VS
DISP
VSC
PT
SETUP
HR
AUTO
ECG
TRACE
SEL
MORE
3. Press the TRACE SEL softkey to display the following menu.
DISP
VSC
DISP
OCRG
TRACE
2
CRG
TRACE
3
AUTO
TRACE
4
AUTO
MORE
4. Press TRACE 2 until CRG is highlighted.
5. Press the PAGE HOME key. OxyCRG information will display in TRACE 2.
To access CRG setup page and available functions:
1. Press softkey next to the CRG label. Available functions are:
SCALE ON softkey to activate parameter scales. Data display is 1.75 minutes if scales
are ON.
CLEAR DATA softkey to clear all stored events from memory.
EVNT SCROL softkey to scroll through stored low alarm events. EVENT 01 is the most
recent.
To record displayed event:
1. Press the RECORD key and within two seconds press softkey next to CRG label. Displayed
EVENT will be recorded to bedside recorder.