1-2
10. Patient Type Selection Key (Adult, Ped, Neo)
11. Start/Stop Manual Recordings
12. Display System Setup Menu
13. Return to Home Page
14. Seven Parameter/System Softkeys
Fixed Function Keys (USA and International)
The Prism & Prism SE monitor includes the following nine (9) fixed functions keys: ON/STBY,
ALARM SUSPEND, ALARM SETUP, FREEZE, NIBP START/STOP, PATIENT TYPE,
SYSTEM SETUP, PAGE HOME, and RECORD. Each of these keys, as the name implies (or
symbol for International), is fixed and performs its intended function regardless of the monitoring
scenario. A brief description of each is provided below.
ON/STBY
Turns the display on and initializes the monitor when ON is pressed. When the monitor is turned
on, you will hear two beeps. The monitor is returned to standby status by pressing STBY. When
in STBY, the batteries continue to be charged if the monitor is connected to AC power, but no
monitoring or storing of data occurs. The Prism SE batteries will also charge when using eternal
DC power.
ALARM SUSPEND
Temporarily suspends alarm tones. The factory defaults for alarm suspend intervals are 180
seconds in adult mode, 90 seconds in pediatric mode, and 60 seconds in neonatal mode. If an
alarm limit is violated, pressing ALARM SUSPEND will silence the alarm tone. Pressing the
key again before the suspend time period has elapsed will reactivate the audible alarm tones if the
alarm condition still exists. While alarms are suspended, the following message will appear on the
monitor screen:
ALARMS SUSPENDED XXX
The XXX will be replaced with the number of seconds remaining until the alarm suspend
condition is removed.